Table 12-10: Step 4 Settings for Release Adherent Cells And Harvest (continued)
Setting
Factory Default
Setting Options
Outlet
EC Outlet
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
Same as step 2
Stop Condition
Time (4.0 min)
Time (0.1 to 20 min)
5.
Confirm the settings for step 5 shown in Table 12-11.
Table 12-11: Step 5 Settings for Release Adherent Cells And Harvest
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
400 mL/min
100 to 500 mL/min
IC Circulation Rate
−70 mL/min
−87 to −18 mL/min
The system automatically calculates this rate based on the IC Inlet Rate
for this step.
EC Inlet
EC Media
Reagent, IC Media, Wash, EC Media
EC Inlet Rate
60 mL/min
0 to 100 mL/min
EC Circulation Rate
30 mL/min
0 to 300 mL/min
Outlet
Harvest
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
In Motion (−180° to 270°, 0 to 3600 sec)
Stop Condition
IC Volume (378 mL)
IC Volume (50 to 1000 mL)
6.
Touch
Start
.
7.
Check the status line to confirm that the task has started.
When the system completes the task, the home screen appears, the system changes to an idle state,
and the status line displays Idle.
8.
Seal and remove the harvest bag.
Release and Harvest Tasks
12-8
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...