Table 12-13: Settings for Harvest Cells (continued)
Setting
Factory Default
Setting Options
EC Circulation Rate
30 mL/min
0 to 300 mL/min
Outlet
Harvest
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
In Motion (−180° to 270°, 0 to 3600 sec)
Stop Condition
IC Volume (378 mL)
IC Volume (50 to 1000 mL)
2.
Touch
Start
.
3.
Check the status line to confirm that the task has started.
When the system completes the task, the home screen appears, the system changes to an idle state,
and the status line displays Idle.
4.
Seal and remove the harvest bag.
Release and Harvest Tasks
12-10
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...