5.
Remove the sample coil strain relief from the sample coil line, as shown in Figure 5-4.
Figure 5-4: Removing the sample coil strain relief
6.
Load the sample coil line 1 to 2 inches from the midpoint into the sterile tubing welder, as shown in
Figure 5-5, and complete the sterile weld with the sterile tubing welder.
Creating a sterile weld at the midpoint of the sample coil may hinder fluid flow when the sample coil
is loaded into the sample coil strain relief.
Figure 5-5: Sample coil loaded into the sterile tubing welder
Note:
Sterile welding near the midpoint of the sample coil allows you to simultaneously take a sample and
reconnect the sample coil.
7.
Set aside the sample you took from the sample coil.
8.
Open the sterile weld on the sample coil.
9.
Connect the sample coil strain relief to the sample coil line.
10.
Recoil the sample coil around the rocker assembly.
11.
Connect the sample coil strain relief to the rocker assembly.
12.
Close the incubator door.
13.
Touch
Resume
.
Operator Procedures
5-6
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...