Coat Bioreactor
The purpose of this task is to coat the bioreactor membrane with a reagent.
•
Step 1: loads a reagent into the IC circulation loop until the bag is empty.
•
Step 2: chases the reagent from the ARC into the IC circulation loop.
•
Step 3: circulates the reagent in the IC circulation loop.
Table 8-4 describes the types of solution that are needed to attach to each line when performing Coat
Bioreactor. These solutions and corresponding volumes are based on the factory default settings for this
task.
As a precondition before starting this task:
Fill the cell inlet bag with a minimum of approximately 40 mL of air.
Table 8-4: Solutions for Coat Bioreactor
Line
Solution Attached to the
Line
Required Volume (estimate based on factory
default settings)
Cell
None
N/A
Reagent
Reagent
As desired
IC Media
None
N/A
Wash
Wash solution
0.1 L plus 6 mL/hr
EC Media
None
N/A
Coat Bioreactor pathway:
Task
>
System Management
>
Coat Bioreactor
1.
Confirm the settings for step 1 shown in Table 8-5.
Table 8-5: Step 1 Settings for Coat Bioreactor
Setting
Factory Default
Setting Options
IC Inlet
Reagent
Cell, Reagent
IC Inlet Rate
10 mL/min
0.1 to 100 mL/min
IC Circulation Rate
100 mL/min
−300 to 300 mL/min
The system automatically calculates this rate based on the IC Inlet Rate
for this step.
EC Inlet
None
N/A
EC Inlet Rate
0 mL/min
N/A
EC Circulation Rate
30 mL/min
0 to 100 mL/min
System Management Tasks
8-8
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...