Remove IC Air
Caution:
Periodically inspect the IC (intracapillary) inlet header and the IC outlet header for air, and if either header
contains visible air, remove the air by performing the Remove IC Air task.
The purpose of this task is to remove any visible air from the inlet port and the outlet port of the IC side
of the bioreactor. Figure 8-1 shows the IC inlet and IC outlet headers located on the bioreactor.
Figure 8-1: IC headers
1
IC inlet header
2
IC outlet header
To perform Remove IC Air:
1.
Touch
Task
.
2.
Touch
System Management
.
3.
Touch
Remove IC Air
.
The Setup: Remove IC Air screen appears, which prompts you to select the inlet source, as shown in
Figure 8-2.
System Management Tasks
8-2
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...