Rapid IC Washout
The purpose of this task is to remove non
‑
adherent cells from the bioreactor. This task imposes
ultrafiltration onto the fiber across the entire fiber length.
Table 9-3 describes the types of solution that are needed to attach to each line when performing Rapid
IC Washout. These solutions and corresponding volumes are based on the factory default settings for
this task.
Table 9-3: Solutions for Rapid IC Washout
Line
Solution Attached to the
Line
Required Volume (estimate based on factory
default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
Media with protein
0.6 L
Wash
None
N/A
EC Media
Media without protein
0.1 L
Rapid IC Washout pathway:
Task
>
Washout
>
Rapid IC Washout
1.
Confirm the settings for Rapid IC Washout shown in Table 9-4.
Table 9-4: Settings for Rapid IC Washout
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
260 mL/min
50 to 500 mL/min
IC Circulation Rate
−45 mL/min
N/A
The system automatically calculates this rate based on the IC Inlet Rate
for this task.
EC Inlet
EC Media
Reagent, IC Media, Wash, EC Media
EC Inlet Rate
40 mL/min
0 to 100 mL/min
EC Circulation Rate
30 mL/min
10 to 300 mL/min
Outlet
IC Outlet
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
In Motion (−180° to 270°, 0 to 3600 sec), Stationary
(−180° to 270°)
Rapid IC Washout
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
9-3
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...