Add Bolus
The purpose of this task is to quickly add a selected volume of reagent into the IC circulation loop; you
can also add a bolus into the EC circulation loop at the same time.
Table 11-5 describes the types of solution that are needed to attach to each line when performing Add
Bolus. These solutions and corresponding volumes are based on the factory default settings for this task.
Table 11-5: Solutions for Add Bolus
Line
Solution
Attached to the
Line
Required Volume (estimate based on factory default settings)
Cell
None
N/A
Reagent
Reagent
0.1 L
IC Media
None
N/A
Wash
None
N/A
EC Media
None
N/A
Add Bolus pathway:
Task
>
Feed and Add
>
Add Bolus
1.
Confirm the settings for Add Bolus shown in Table 11-6.
Table 11-6: Settings for Add Bolus
Setting
Factory Default
Setting Options
IC Inlet
Reagent
Reagent, IC Media, Wash, EC Media, None
IC Inlet Rate
10 mL/min
0 to 200 mL/min
IC Circulation Rate
100 mL/min
−300 to 300 mL/min
The system automatically calculates this rate based on the IC Inlet Rate
for this task.
EC Inlet
None
Reagent, IC Media, Wash, EC Media, None
EC Inlet Rate
0 mL/min
0 to 300 mL/min
EC Circulation Rate
30 mL/min
0 to 300 mL/min
Outlet
EC Outlet
IC Outlet, EC Outlet, Harvest
Rocker
Stationary (0°)
In Motion (−180° to 270°, 0 to 3600 sec), Stationary
(−180° to 270°)
Feed and Add Tasks
11-6
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...