Attach Cells
The purpose of this task is to allow adherent cells to attach to the bioreactor membrane while allowing
flow in the EC circulation loop. The pump flow rate to the IC circulation loop is set to zero.
Table 10-8 describes the types of solution that are needed to attach to each line when performing Attach
Cells. These solutions and corresponding volumes are based on the factory default settings for this task.
Table 10-8: Solutions for Attach Cells
Line
Solution Attached to the
Line
Required Volume (estimate based on factory
default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
None
N/A
Wash
None
N/A
EC Media
Media without protein
6 mL/hour
Attach Cells pathway:
Task
>
Load and Attach
>
Attach Cells
1.
Confirm the settings for Attach Cells shown in Table 10-9.
Table 10-9: Settings for Attach Cells
Setting
Factory Default
Setting Options
IC Inlet
None
N/A
IC Inlet Rate
0 mL/min
N/A
IC Circulation Rate
0 mL/min
N/A
EC Inlet
EC Media
Reagent, IC Media, Wash, EC Media
EC Inlet Rate
0.1 mL/min
0.1 to 10 mL/min
EC Circulation Rate
30 mL/min
0 to 100 mL/min
Outlet
EC Outlet
N/A
Rocker
Stationary (0°)
Stationary (0° to 180°)
Stop Condition
Manual
Time (0.1 to 2880 min), Manual, EC Volume
(1 to 4000 mL)
2.
Touch
Start
.
3.
Check the status line to confirm that the task has started.
4.
Touch
Finish
when you are ready to complete the task.
Attach Cells
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
10-7
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...