
Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
298
06807A(DCNxxxx)
20. Perform a leak check the system.
21. Power up the analyzer and verify the basic operation of the analyzer using the
ETEST and OTEST features (refer to Section 6.9.5 and 6.9.6) or by measuring
calibrated zero and span gases.
22. Allow the instrument to warm up for 60 minutes.
23. Perform a PMT Hardware calibration (refer to Section 10.7.2.8).
24. Perform a zero point and span calibration (refer to Section 6).
10.7.2.8.
PMT HARDWARE CALIBRATION (
FACTORY CAL
)
The sensor module hardware calibration adjusts the slope of the PMT output when the
instrument’s slope and offset values are outside of the acceptable range and all other more
obvious causes for this problem have been eliminated.
Figure 10-17: Pre-Amplifier Board (Preamp PCA) Layout
1. Set the instrument reporting range type to
SNGL
(refer to Section 4.6.4).
2. Perform a zero
–point calibration using zero air (refer to Section 6).
3. Let the instrument stabilize by allowing it to run for one hour.
4. Adjust the UV Lamp (refer to Section 10.7.2.5).
5. Perform a
LAMP CALIBRATION
procedure (refer to Section 4.8.7).
6. Locate the Preamp PCA (refer to Figure 10-16).
7. Locate the Following Components On the Preamp PCA (Figure 10-17):
8. HVPS coarse adjustment switch (Range 0-9, then A-F).
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...