
Theory Of Operation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
258
06807A(DCNxxxx)
which allows the instrument to compensation of the SO
2
calculations based on changes
in ambient pressure.
When the TPC feature is enabled, the analyzer’s SO
2
concentration divided by a factor
call PRESSCO which is based on the difference between the ambient pressure of the
sample gas normalized to standard atmospheric pressure (Equation 9-6). As ambient
pressure increases, the compensated SO
2
concentration is decreased.
A)
-
HG
("
29.92
_SLOPE
SAMP_PRESS
A)
-
HG
("
SSURE
SAMPLE_PRE
PRESSCO
(Equation 9-6)
SAMPLE-PRESSURE
: The ambient pressure of the sample gas as measured by the
instrument’s sample pressure sensor in “Hg-A.
SAMP_PRESS_SLOPE
: Sample pressure slope correction factor. The default setting
for Section 6.8 describes the method for enabling/disabling the TPC feature.
9.7.4. INTERNAL DATA ACQUISITION SYSTEM (DAS)
The DAS is designed to implement predictive diagnostics that stores trending data for
users to anticipate when an instrument will require service. Large amounts of data can
be stored in non-volatile memory and retrieved in plain text format for further
processing with common data analysis programs. The DAS has a consistent user
interface in all Teledyne ML instruments. New data parameters and triggering events
can be added to the instrument as needed.
Depending on the sampling frequency and the number of data parameters the DAS can
store several months of data, which are retained even when the instrument is powered
off or a new firmware is installed. The DAS permits users to access the data through the
instrument’s front panel or the remote interface. The latter can automatically download
stored data for further processing. For information on using the DAS, refer to Section
5.2..
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...