Operating Instructions
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
80
The Mode field of the front panel display indicates to the user which operating mode the
unit is currently running.
Besides
SAMPLE
and
SETUP
, other modes the analyzer can be operated in are:
Table 4-1: Analyzer Operating Modes
MODE
EXPLANATION
DIAG
One of the analyzer’s diagnostic modes is active (refer to Section 4.8).
LO CAL A
Unit is performing LOW SPAN
(midpoint) calibration initiated automatically by the analyzer’s
AUTOCAL feature
LO CAL R
Unit is performing LOW SPAN (midpoint) calibration initiated remotely through the COM ports or
digital control inputs.
M-P CAL
This is the basic calibration mode of the instrument and is activated by pressing the CAL button.
SAMPLE
Sampling normally, flashing text indicates adaptive filter is on.
SAMPLE A
Indicates that unit is in SAMPLE mode and AUTOCAL feature is activated.
SETUP
SETUP mode is being used to configure the analyzer. The gas measurement will continue during
this process.
SPAN CAL A
1
Unit is performing SPAN calibration initiated automatically by the analyzer’s AUTOCAL feature
SPAN CAL M
1
Unit is performing SPAN calibration initiated manually by the user.
SPAN CAL R
1
Unit is performing SPAN calibration initiated remotely through the COM ports or digital control
inputs.
ZERO CAL A
1
Unit is performing ZERO calibration procedure initiated automatically by the AUTOCAL feature
ZERO CAL M
1
Unit is performing ZERO calibration procedure initiated manually by the user.
ZERO CAL R
1
Unit is performing ZERO calibration procedure initiated remotely through the COM ports or digital
control inputs.
1
Only Appears on units with Z/S valve or IZS options..
Finally, the various
CAL
modes are for calibrating the analyzer. This mode is described
separately in Section 6.
4.2. SAMPLE MODE
This is the analyzer’s standard operating mode. In this mode, the instrument is analyzing
SO
2
and calculating concentrations.
4.2.1. TEST FUNCTIONS
A series of test functions is available at the front panel while the analyzer is in
SAMPLE
mode. These parameters provide information about the present operating
status of the instrument and are useful during troubleshooting (refer to Section 10.1.2).
They can also be recorded in one of the DAS channels (refer to Section 5.2) for data
analysis. To view the test functions, press one of the
<TST TST>
buttons repeatedly in
either direction.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...