
EPA Protocol Calibration
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
200
06807A(DCNxxxx)
7.1.5. CALIBRATION FREQUENCY
To ensure accurate measurements of the SO
2
concentrations, calibrate the analyzer at the
time of installation, and re-calibrate it:
No later than three months after the most recent calibration or performance audit to indicate
an acceptable analyzer calibration.
An interruption of more than a few days in analyzer operation.
Any repairs which might affect its calibration.
Physical relocation of the analyzer.
Any other indication (including excessive zero or span drift) of possible significant
inaccuracy of the analyzer.
Following any of the activities listed above, the zero and span should be checked to
determine if a calibration is necessary. If the analyzer zero and span drifts exceed locally
established calibration units or the calibration limits in Section 2.0.9, Subsection 9.1.3
(Q.A. Handbook), a calibration should be performed.
7.1.6. RECORD KEEPING
Record keeping is a critical part of all quality assurance programs. Standard forms
similar to those that appear in this manual should be developed for individual programs.
Three things to consider in the development of record forms are:
Does the form serve a necessary function?
Is the documentation complete?
Will the forms be filed in such a manner that they can easily be retrieved when needed?
7.1.7. SUMMARY OF QUALITY ASSURANCE CHECKS
The following items should be checked on a regularly scheduled basis to assure high
quality data from the T100. Refer to Table 7-3 for a summary of activities. Also the QA
Handbook should be checked for specific procedures.
Table 7-3: Activity Matrix for Quality Assurance Checks
CHARACTERISTIC
ACCEPTANCE LIMITS
FREQUENCY AND METHOD OF
MEASUREMENT
ACTION IF REQUIREMENTS ARE
NOT MET
Shelter temperature
Mean temperature between
22
o
C and 28
o
C (72
o
and
82
o
F), daily fluctuations not
greater than ±2
o
C
Check thermograph chart
weekly for variations greater
than ±2
o
C (4
o
F)
Mark strip chart for the affected
time period
Repair or adjust temperature
control
Sample introduction
system
No moisture, foreign
material, leaks, obstructions;
sample line connected to
manifold
Weekly visual inspection
Clean, repair, or replace as
needed
Recorder
Adequate ink & paper
Legible ink traces
Correct chart speed and
range
Correct time
Weekly visual inspection
Replenish ink and paper supply
Adjust time to agree with clock;
note on chart
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...