Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
174
OR
SAMPLE
EXHAUST
MODEL 701
Zero Air
Generator
VE
N
T
Calibrated
SO
2
GAS
(At high
concentration)
MODEL 700
Gas Dilution
Calibrator
(with Ozone
Bench Option)
Source of
SAMPLE Gas
(Remove
during
calibration)
SAMPLE
EXHAUST
VE
N
T
Calibrated
SO
2
GAS
(At high
concentration)
MODEL 701
Zero Air
Generator
Source of
SAMPLE Gas
(Remove
during
calibration)
3-way
Valve
Needle valve
to control
flow
Chassis
Chassis
Figure 6-1: Setup for Manual Calibration without Z/S valve or IZS Option (Step 1)
STEP TWO
: Set the expected
SO
2
span gas concentrations. In this example the
instrument is set for single (
SNGL
) range mode with a reporting range span of 500 ppb.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...