Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Remote Operation & Advanced Features
127
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SETUP X.
MACHINE ID: 100 ID
0 1 0 0
ENTR EXIT
Toggle these keys
to
cycle through the
available character set:
0-9
ENTR
key accepts the
new settings
EXIT
key ignores the
new settings
Figure 5-2: COMM
– Analyzer ID
The ID can be any 4-digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)
5.1.2. ETHERNET CONFIGURATION
When using the Ethernet interface, the analyzer can be connected to any standard
10BaseT or 100BaseT Ethernet network via low-cost network hubs, switches or routers.
The interface operates as a standard TCP/IP device on port 3000. This allows a remote
computer to connect through the network to the analyzer using APICOM, terminal
emulators or other programs.
The Ethernet cable connector on the rear panel has two LEDs indicating the Ethernet’s
current operating status.
Table 5-1: Ethernet Status Indicators
LED
FUNCTION
amber (link)
On when connection to the LAN is valid.
green (activity
Flickers during any activity on the LAN.
The analyzer is shipped with DHCP enabled by default. This allows the instrument to be
connected to a network or router with a DHCP server. The instrument will automatically
be assigned an IP address by the DHCP server (Section 5.1.2.2). This configuration is
useful for quickly getting an instrument up and running on a network. However, for
permanent Ethernet connections, a static IP address should be used. Section 5.1.2.1
below details how to configure the instrument with a static IP address.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...