Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Calibration Procedures
191
Press EXIT 3
times to return
to SAMPLE
menu.
SETUP X.X
STABIL_GAS:O2
SO2 CO2 O2
ENTR
EXIT
SAMPLE RANGE=50.0 PPM SO2= XX.XX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG ALRM
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
2) STABIL_GAS=SO2
<PREV NEXT> JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS:SO2
SO2 CO2
O2
ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
CO2 and O2
options only
appear if
associated
sensors are
installed.
Figure 6-15:
Activate O
2
Sensor Stability Function
IMPORTANT
Use the same procedure to reset the STB test function to SO
2
when the O
2
calibration
procedure is complete
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...