Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
172
Devices such as the API Model 701 zero air generator that condition ambient air by
drying and removal of pollutants are available. We recommend this type of device for
generating zero air.
6.1.1.2.
SPAN GAS
Span gas is specifically mixed to match the chemical composition of the gas being
measured at about 80% of the desired full measurement range. For example, if the
measurement range is 500 ppb, the span gas should have an SO
2
concentration of about
400 ppb.
Span gases should be certified to a specific accuracy to ensure accurate calibration of the
analyzer. Typical gas accuracy for SO
2
gases is 1 or 2 %. SO
2
standards should be mixed
in nitrogen.
Teledyne ML offers an IZS option operating with permeation devices. The accuracy of
these devices is about ±5%. Whereas this may be sufficient for quick, daily calibration
checks, we strongly recommend using certified SO
2
span gases for accurate calibration.
IMPORTANT
Applications requiring US-EPA equivalency do not allow permeation devices to be used
as sources of span gas for calibration of the analyzer.
6.1.1.3.
CALIBRATION GAS STANDARDS AND TRACEABILITY
All equipment used to produce calibration gases should be verified against standards of
the National Institute for Standards and Technology (NIST). To ensure NIST
traceability, we recommend acquiring cylinders of working gas that are certified to be
traceable to NIST Standard Reference Materials (SRM). These are available from a variety
of commercial sources.
Table 6-1: NIST-SRM's Available for Traceability of SO
2
Calibration Gases
NIST-SRM
4
TYPE
NOMINAL
CONCENTRATION
1693a
1694a
1661a
Sulfur dioxide in N
2
Sulfur dioxide in N
2
Sulfur dioxide in N
2
50 ppm
100 ppm
500 ppm
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...