
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Troubleshooting & Repair
06807A(DCNxxxx)
283
Check for the physical presence of the valves or the IZS option.
Check that a working perm-tube is installed in the IZS oven assembly.
Check front panel for correct software configuration. When the instrument is in
SAMPLE mode, the front panel display should show CALS and CALZ buttons in the
second line of the display. The presence of the buttons indicates that the option has
been enabled in software. In addition, the IZS option is enabled if the TEST functions
show a parameter named IZS TEMP.
The IZS option is heated with a proportional heater circuit and the temperature is
maintained at 50° C ±1°. Check the
IZS TEMP
function via front panel display (refer to
Section 4.2.1) and the
IZS_TEMP
signal voltage using the
SIGNAL I/O
function under
the
DIAG
Menu (refer to Section 4.8.2).
At 50°
C, the temperature signal from the IZS thermistor should be around 2500 mV.
10.6.18. BOX TEMPERATURE
The box temperature sensor (thermistor) is mounted on the motherboard at the bottom,
right corner of the CPU board when looking at it from the front. It cannot be disconnected
to check its resistance. Box temperature will vary with, but will always read about 5° C
higher than, ambient (room) temperature because of the internal heating zones sample
chamber and other devices.
To check the box temperature functionality, we recommend checking the
BOX_TEMP
signal voltage using the
SIGNAL I/O
function under the
DIAG
Menu (refer to Section
At about 30°
C (5
above typical room temperature), the signal should be around 1500
mV. We recommend using a certified or calibrated external thermometer / temperature
sensor to verify the accuracy of the box temperature.
10.6.19. PMT TEMPERATURE
PMT temperature should be low and constant. It is more important that this temperature is
maintained constant than it is to maintain it low. The PMT cooler uses a Peltier, thermo-
electric element powered by 12 VDC from the switching power supply PS2. The
temperature is controlled by a proportional temperature controller located on the
preamplifier board. Voltages applied to the cooler element vary from +/- 0.1 to +/- 12
VDC. The temperature set point (hard-wired into the preamplifier board) will vary by
about ±1
C due to component tolerances. The actual temperature will be maintained to
within 0.1
C around that set point.
On power-up of the analyzer, the front panel enables the user to watch that temperature
drop from about ambient temperature down to its set point of 6-8° C.
If the temperature fails to drop after 20 minutes, there is a problem in the cooler circuit.
If the control circuit on the preamplifier board is faulty, a temperature of -1
C is
reported.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...