Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
111
4.8.4.3.
MANUAL ANALOG OUTPUT CALIBRATION AND VOLTAGE ADJUSTMENT
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Calibration is done through the instrument software with a voltmeter connected across
the output terminals (refer to Figure 4-24). Adjustments are made using the control
buttons by setting the zero-point first and then the span-point (refer to Table 4-12).
The software allows this adjustment to be made in 100, 10 or 1 count increments.
Table 4-12: Voltage Tolerances for Analog Output Calibration
FULL SCALE
ZERO TOLERANCE
SPAN VOLTAGE
SPAN TOLERANCE
0.1 VDC
±0.0005V
90 mV
±0.001V
1 VDC
±0.001V
900 mV
±0.001V
5 VDC
±0.002V
4500 mV
±0.003V
10 VDC
±0.004V
4500 mV
±0.006V
IMPORTANT
Outputs configured for 0.1V full scale should always be calibrated manually.
V
+DC Gnd
V OUT +
V OUT -
V IN +
V IN -
Recording
Device
ANALYZER
See Table 3-1 for
pin assignments
of Analog Out
connector on the
rear panel
Figure 4-24:
Setup for Calibrating Analog Outputs
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...