Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
262
06807A(DCNxxxx)
To view or clear a warning message press:
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
TEST
CAL
MSG
CLR
SETUP
If warning messages re-appear,
the cause needs to be found. Do
not repeatedly clear warnings
without corrective action.
Press
CLR
to clear the current
warning message.
If more than one warning is
active, the next message will
take its place.
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
Mode.
SAMPLE
RANGE = 500.0 PPB
SO2 =XXX.X
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
SYSTEM RESET
SO2= X.XXX
< TST TST > CAL
MSG
CLR
SETUP
In
WARNING
mode,
<TST TST
>
keys replaced with
TEST
key.
Pressing
TEST
switches to
SAMPLE mode and hides warning
messages until new warning(s)
are activated.
MSG
indicates that one or more
warning message are active but
hidden. Pressing
MSG
cycles
through warnings
In
SAMPLE
mode, all warning
messages are hidden, but
MSG
button appears
Figure 10-1:
Viewing and Clearing Warning Messages
Table 10-1: Warning Messages - Indicated Failures
Warning Message
Fault Condition
Possible Causes
ANALOG CAL
WARNING
The instruments A/D
circuitry or one of its analog
outputs is not calibrated
A parameter for one of the analog outputs, even one not currently being used,
has been changed and the analog output calibration routine was not re-run
A/D circuitry failure on motherboard
Other motherboard electronic failure
BOX TEMP WARNING
Box Temp is < 5°C or >
48°C.
NOTE: Box temperature typically runs ~7
o
c warmer than ambient temperature.
Poor/blocked ventilation to the analyzer.
Stopped exhaust-fan
Ambient temperature outside of specified range
CANNOT DYN SPAN
Dynamic Span operation
failed
Measured concentration value is too high or low.
Concentration slope value too high or too low
CANNOT DYN ZERO
Dynamic Zero operation
failed
Measured concentration value is too high.
Concentration offset value too high.
CONFIG INITIALIZED
Configuration and
Calibration data reset to
original Factory state.
Failed disk on module
User erased data
DARK CAL WARNING
The Dark Cal signal is
higher than 200 mV.
Light leak in reaction cell
Shutter solenoid is not functioning
Failed relay board
I
2
C bus failure
Loose connector/wiring
PMT preamp board bad or out of cal
DATA INITIALIZED
Data Storage in DAS was
erased
Failed disk on module
User cleared data
HVPS WARNING
High voltage power supply
output is <400 V or >900 V
High voltage power supply is bad
High voltage power supply is out of cal
A/D converter circuitry is bad
IZS TEMP WARNING
On units with IZS options
installed: The permeation
tube temperature is Sample
chamber temperature is
< 45°C or > 55°C
Bad IZS heater
Bad IZS temperature sensor
Bad relay controlling the IZS heater
Entire relay board is malfunctioning
I
2
C bus malfunction
Failure of thermistor interface circuitry on motherboard
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...