Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
164
The Hessen protocol is not well defined; therefore while ML’s application is completely
compatible with the protocol itself, it may be different from implementations by other
companies.
The following subsections describe the basics for setting up your instrument to operate
over a Hessen Protocol network. For more detailed information as well as a list of host
computer commands and examples of command and response message syntax,
download the
Manual Addendum for Hessen Protocol
from the Teledyne ML’s website:
http://www.teledyne-ml.com/manuals/index.asp
.
5.3.4.2.
HESSEN COMM PORT CONFIGURATION
Hessen protocol requires the communication parameters of the T100’s COMM ports to
be set differently than the standard configuration as shown in the table below.
Table 5-13: RS-232 Communication Parameters for Hessen Protocol
Parameter
Standard
Hessen
Data Bits
8
7
Stop Bits
1
2
Parity
None
Even
Duplex
Full
Half
To change the rest of the COMM port parameters and modes, refer to Section 5.1.4.
To change the baud rate of the T100’s COMM ports, refer to Section 5.1.5.
IMPORTANT
Ensure that the communication parameters of the host computer are also properly set.
NOTE
The instrument software has a 200 ms latency before it responds to commands issued by
the host computer. This latency should present no problems, but you should be aware of
it and not issue commands to the instrument too quickly.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...