Operating Instructions
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
90
In order to compensate for CPU clocks which run fast or slow, there is a variable to
speed up or slow down the clock by a fixed amount every day. To change this variable,
press:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS: 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
3x
EXIT
returns
to the main
SAMPLE
display
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 0 ) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
7) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP X.X
CLOCK_ADJ:
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
Continue to press
NEXT
until …
SAMPLE
ENTER SETUP PASS: 818
8 1 8
ENTR
EXIT
Figure 4-8: SETUP
– Clock Speed Variable
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...