Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
150
5.2.2.7.
RS-232 REPORT FUNCTION
The M DAS can automatically report data to the communications ports, where they can
be captured with a terminal emulation program or simply viewed by the user.
To enable automatic COMM port reporting, follow the instruction shown in Figure 5-11,
then press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
RS-232 REPORT: OFF
<SET SET>
EDIT
EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
RS-232 REPORT: OFF
OFF
ENTR EXIT
Toggle to turn
reporting ON or OFF
Press
SET>
until…
Figure 5-18:
DAS
– RS-232 Report Function
5.2.2.8.
COMPACT REPORT
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead
of reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line, instead.
5.2.2.9.
STARTING DATE
This option allows a user to specify a starting date for any given channel in case the user
wants to start data acquisition only after a certain time and date. If the
Starting Date
is
in the past, the DAS ignores this setting.
5.2.2.10.
DISABLING/ENABLING DATA CHANNELS
Data channels can be temporarily disabled, which can reduce the read/write wear on the
disk-on-module. The ALL_01 channel of the T100, for example, is disabled by default.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...