Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
57
IMPORTANT
EPA Requirements:
US EPA requirements state that zero air and span gases be supplied at twice the
instrument’s specified gas flow rate. Therefore for the T100 zero and span gases should
be supplied to their respective inlets in excess of 1300 cc
3
/min (650 cc
3
/min. x 2).
NOTE
Dust Plugs:
In order to prevent dust from getting into the gas flow channels of your analyzer, it was
shipped with small plugs inserted into each of the pneumatic fittings on the back panel.
Ensure that all of these dust plugs are removed before attaching exhaust and supply gas
lines.
IMPORTANT
Leak Check:
Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using the procedures defined in Section 8.3.6.
CAUTION
GENERAL SAFETY HAZARD
Gas flow though the analyzer must be maintained at all time for units with a permeation
tube installed. Insufficient gas flow allows gas to build up to levels that will contaminate
the instrument or present a safety hazard to personnel.
Section 3.5.1 provides pneumatic connection instructions, and Table 3-9 provides
reference to the location of various pneumatic layout configurations.
Table 3-9: Pneumatic Layout Reference
Pneumatic Layout
Section
Basic
Zero/Span Valves
Internal Zero/Span (IZS)
Basic with O2 Sensor
Basic with CO2 Sensor
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...