Instrument Maintenance
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
220
06807A(DCNxxxx)
2.
Disconnect the pneumatic tubing attached to both ends of the scrubber’s inner
tubing.
3. One end is connected to the sample particulate filter assembly and the other end is
connected to the reaction cell assembly.
4. Both ends are made of the 1/8" black Teflon tubing.
5. Cap one end of the hydrocarbon scrubber.
6. Attach the pressurized air source to the other end of the scrubber inner tubing with
the leak check fixture in line.
Pump
or
Pressurized Air
Source
Leak Check
Fixture
Scrubber
Cap
Figure 8-4: Hydrocarbon Scrubber Leak check Setup
7. Use the needle valve to adjust the air input until the gauge reads 15 psiA.
ATTENTION
COULD DAMAGE INSTRUMENT
Do not exceed a pressure of more than 15 psia.
Do not pull the vacuum through the scrubber.
8. Close the shut-off valve.
9. Wait 5 minutes.
If the gauge pressure drops >1 psi within 5 minutes, then the hydrocarbon scrubber has
an internal leak and must be replaced. Contact Teledyne ML’s Customer Service.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...