Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-29
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
Rear board primary MUX analog inputs
PMT_SIGNAL 0
PMT
detector
HVPS_VOLTAGE
1
HV power supply output
PMT_TEMP 2
PMT
temperature
UVLAMP_SIGNAL
3
UV lamp intensity
4
Temperature
MUX
PHOTO_ABS
8
5
Pre-amplified UV lamp intensity
O2_SENSOR
10
6
O
2
concentration sensor
SAMPLE_PRESSURE 7
Sample
pressure
TEST_INPUT_8
8
Diagnostic test input
REF_4096_MV
9
4.096V reference from MAX6241
SAMPLE_FLOW 10
Sample flow rate
VACUUM_PRESSURE
2
10
Vacuum
pressure
CO2_SENSOR
11
11
CO
2
concentration sensor
12–13
Spare (thermocouple input?)
14
DAC
MUX
REF_GND 15
Ground
reference
Rear board temperature MUX analog inputs
BOX_TEMP
0
Internal box temperature
RCELL_TEMP
1
Reaction cell temperature
IZS_TEMP 2
IZS
temperature
3
Spare
O2_CELL_TEMP
10
4
O
2
sensor cell temperature
TEMP_INPUT_5
5
Diagnostic temperature input
TEMP_INPUT_6
6
Diagnostic temperature input
7
Spare
Rear board DAC MUX analog inputs
DAC_CHAN_1
0
DAC channel 0 loopback
DAC_CHAN_2
1
DAC channel 1 loopback
DAC_CHAN_3
2
DAC channel 2 loopback
DAC_CHAN_4
3
DAC channel 3 loopback
Rear board analog outputs
CONC_OUT_1,
DATA_OUT_1
0
Concentration output #1 (SO
2
, range #1),
Data output #1
CONC_OUT_2,
DATA_OUT_2
1
Concentration output #2 (SO
2
, range #2),
Data output #2
CONC_OUT_3
10,
DATA_OUT_3
2
Concentration output #3 (CO
2
or O
2
),
Data output #3
TEST_OUTPUT,
DATA_OUT_4
3
Test measurement output,
Data output #4
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...