Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Troubleshooting & Repair
06807A(DCNxxxx)
269
To check the actual sample flow, disconnect the sample tube from the sample inlet on the
rear panel of the instrument. Ensure that the unit is in basic SAMPLE mode. Place a
finger over the inlet and see if it gets sucked in by the vacuum or, more properly, use a
flow meter to measure the actual flow. If a proper flow of approximately 650 cm³/min
exists, contact customer service. If there is no flow or low flow, continue with the next
step.
Check that the sample pressure is at or around 28 (or about1 in-Hg-A below ambient
atmospheric pressure).
10.3.2. HIGH FLOW
Flows that are significantly higher than the allowed operating range (typically ±10-11%
of the nominal flow) should not occur in the M unless a pressurized sample, zero or span
gas is supplied to the inlet ports. Be sure to vent excess pressure and flow just before the
analyzer inlet ports.
When supplying sample, zero or span gas at ambient pressure, a high flow would indicate
that one or more of the critical flow orifices are physically broken (very unlikely case),
allowing more than nominal flow, or were replaced with an orifice of wrong
specifications. If the flows are more than 15% higher than normal, we recommend that
the technician find and correct the cause of the flow problem,
10.4. CALIBRATION PROBLEMS
This section provides information regarding possible causes of various calibration
problems.
10.4.1. NEGATIVE CONCENTRATIONS
Negative concentration values may be caused due to the following:
A slight, negative signal is normal when the analyzer is operating under zero gas and the
signal is drifting around the zero calibration point. This is caused by the analyzer’s zero noise
and may cause reported concentrations to be negative for a few seconds at a time down to -5
ppb, but should alternate with similarly high, positive values.
Mis-calibration is the most likely explanation for negative concentration values. If the zero air
contained some SO
2
gas (contaminated zero air or a worn-out zero air scrubber) and the
analyzer was calibrated to that concentration as “zero”, the analyzer may report negative
values when measuring air that contains little or no SO
2
. The same problem occurs, if the
analyzer was zero-calibrated using ambient air or span gas.
If the response offset test function for SO
2
(
OFFSET
) are greater than 150 mV, a failed PMT
or high voltage supply, or sample chamber contamination, could be the cause.
10.4.2. NO RESPONSE
If the instrument shows no response (display value is near zero) even though sample gas
is supplied properly and the instrument seems to perform correctly,
Confirm response by supplying SO
2
span gas of about 80% of the range value to the analyzer.
Check the sample flow rate for proper value.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...