Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
278
06807A(DCNxxxx)
10.6.7.4.
CONTROL INPUTS
The control input bits can be tested by the following procedure:
1. Connect a jumper from the +5 V pin on the STATUS connector to the U on the
CONTROL IN connector.
2. Connect a second jumper from the
pin on the STATUS connector to the A pin on
the CONTROL IN connector. The instrument should switch from
SAMPLE
mode to
ZERO CAL R
mode.
3. Connect a second jumper from the
pin on the STATUS connector to the B pin on
the CONTROL IN connector. The instrument should switch from
SAMPLE
mode to
SPAN CAL R
mode.
In each case, the T100 should return to SAMPLE mode when the jumper is removed.
10.6.8. CPU
There are two major types of CPU board failures, a complete failure and a failure
associated with the Disk-On-Module (DOM). If either of these failures occurs, contact the
factory.
For complete failures, assuming that the power supplies are operating properly and the
wiring is intact, the CPU is faulty if on power-on, the watchdog LED on the motherboard
is not flashing.
In some rare circumstances, this failure may be caused by a bad IC on the motherboard,
specifically U57, the large, 44 pin device on the lower right hand side of the board. If this
is true, removing U57 from its socket will allow the instrument to start up but the
measurements will be invalid.
If the analyzer stops during initialization (the front panel display shows a fault or warning
message), it is likely that the DOM, the firmware or the configuration and data files have
been corrupted.
10.6.9. RS-232 COMMUNICATION
This section provides general RS-232 communication information.
10.6.9.1.
GENERAL RS-232 TROUBLESHOOTING
Teledyne ML’s analyzers use the RS-232 protocol as the standard, serial communications
protocol. RS-232 is a versatile standard, which has been used for many years but, at
times, is difficult to configure. Teledyne ML conforms to the standard pin assignments in
the implementation of RS-232. Problems with RS-232 connections usually center around
4 general areas:
Incorrect cabling and connectors. This is the most common problem. Refer to Section
3.4.8.3 for connector, pin-out and setup information.
The communications (baud) rate and protocol parameters are incorrectly configured.
Refer to Sections 3.4.8.3 and 5.1.5 for baud rate information.
The COMM port communications mode is set incorrectly (refer to Section 5.1.4).
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...