3.4. ELECTRICAL CONNECTIONS
Note
To maintain compliance with EMC standards, it is required that the cable length be no
greater than 3 meters for all I/O connections, which include Analog In, Analog Out, Status
Out, Control In, Ethernet/LAN, USB, RS-232, and RS-485.
This section provides instructions for basic connections and for options. Table 3-4
provides a direct link to the instructions for the subsections that apply to your analyzer’s
configuration.
Table 3-4: Electrical Connections References
Connection
Basic Option Section
Power
X
Analog Input
X
Analog Output
X
Current Loop Analog Outputs
X
Status Outputs
X
Control Inputs
X
Concentration Alarm Relay
X
Ethernet
X
USB COM Port
X
RS-232
X
RS-232 Multidrop
X
RS-485
X
3.4.1. CONNECTING POWER
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying
at least 10 Amps of current at your AC voltage and that it is equipped with a functioning
earth ground.
WARNING
ELECTRICAL SHOCK HAZARD
High Voltages are present inside the analyzers case.
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Power off analyzer before disconnecting or connecting electrical subassemblies.
Do not operate analyzer with the cover off.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...