Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
148
To define the
REPORT PERIOD
, follow the instruction shown in Figure 5-11, then
press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu.
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
Press
SET>
until you reach
REPORT PERIOD
…
SETUP X.X
REPORT PERIODD:DAYS:0
0 0 0
ENTR EXIT
ENTR
accepts the new string and
returns to the previous menu.
EXIT
ignores the new string and
returns to the previous menu.
Press buttons to set amount of
time between reports, in hours
(HH) and/or minutes (MM)
(max: 23:59). 01:00 sets a
report to be made every hour.
Set the number of days
between reports (0-365).
IIf at any time an invalidl entry is selected (e.g., days > 366)
the ENTR button will disappear from the display.
SETUP X.X
REPORT PERIODD:TIME:01:00
0 1 0 0
ENTR EXIT
SETUP X.X
REPORT PERIOD:000:00:05
<SET SET>
EDIT
EXIT
Use PREV and NEXT to
scroll to the data channel
to be edited.
Figure 5-16:
DAS
– Define the Report Period
5.2.2.6.
NUMBER OF RECORDS
The number of data records in the T100 is limited to about a cumulative one million data
points in all channels (one megabyte of space on the disk-on-module). However, the
actual number of records is also limited by the total number of parameters and channels
and other settings in the DAS configuration. Every additional data channel, parameter,
number of samples setting etc. will reduce the maximum amount of data points
somewhat. In general, however, the maximum data capacity is divided amongst all
channels (max: 20) and parameters (max: 50 per channel).
The DAS will check the amount of available data space and prevent the user from
specifying too many records at any given point. If, for example, the DAS memory space
can accommodate 375 more data records, the ENTR button will disappear when trying
to specify more than that number of records. This check for memory space may also
make an upload of a DAS configuration with APICOM or a Terminal program fail, if
the combined number of records would be exceeded. In this case, it is suggested to
either try from the front panel what the maximum number of records can be or use trial-
and-error in designing the DAS script or calculate the number of records using the DAS
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...