APPENDIX A - Version Specific Software Documentation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
A-28
05036 Rev D
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
Relay board digital output (PCF8575), default I
2
C address 44 hex
RELAY_WATCHDOG 0
Alternate between 0 and 1 at least every 5 seconds to keep
relay board active
RCELL_HEATER
1
0 = reaction cell heater on
1 = off
2–3
Spare
IZS_HEATER
4
0 = IZS heater on
1 = off
O2_CELL_HEATER
10
5
0 = O
2
sensor cell heater on
1 = off
CAL_VALVE
6
0 = let cal. gas in
1 = let sample gas in
SPAN_VALVE
7
0 = let span gas in
1 = let zero gas in
LOW_SPAN_VALVE
2, 6
8
0 = let low span gas in
1 = let sample gas in
CYLINDER_VALVE
7
8
0 = open pressurized inlet valve
1 = close valve
ZERO_VALVE
2
9
0 = let zero gas in
1 = let sample gas in
DARK_SHUTTER
10
0 = close dark shutter
1 = open
11–15
Spare
AUX board analog inputs, default I
2
C address 30 hex
PMT_SIGNAL
3
0 (register number)
PMT detector
UVLAMP_SIGNAL
3
1
UV lamp intensity
NORM_PMT_SIGNAL
3
2
Normalized PMT detector
PMT_TEMP
3
3
PMT
temperature
HVPS_VOLTAGE
3
4
HV power supply output
PMT_DARK
3
5
PMT reading during dark cycles
LAMP_DARK
3
6
Lamp reading during dark cycles
AGND_DARK
3
7
AGND reading during dark cycles
AGND_LIGHT
3
8
AGND reading during light cycles
VREF_DARK
3
9
VREF4096 reading during dark cycles
VREF_LIGHT
3
10
VREF4096 reading during light cycles
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...