Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-17
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
BAUD_RATE2 —
19200
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM2 baud rate. Enclose
value in double quotes (“) when setting
from the RS-232 interface.
MODEM_INIT2 —
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0 E0
Q1 &W0”
Any character in
the allowed
character set. Up to
100 characters
long.
RS-232 COM2 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
RS232_PASS Password
940331
0–999999
RS-232 log on password.
MACHINE_ID
ID
100
0–9999
Unique ID number for instrument.
COMMAND_PROMPT —
“Cmd>
” Any character in
the allowed
character set. Up to
100 characters
long.
RS-232 interface command prompt.
Displayed only if enabled with
RS232_MODE
variable. Enclose value
in double quotes (“) when setting from
the RS-232 interface.
TEST_CHAN_ID —
NONE
NONE,
PMT READING,
UV READING,
VACUUM
PRESSURE
3
,
SAMPLE
PRESSURE,
SAMPLE FLOW,
RCELL TEMP,
O2 CELL TEMP
10
,
CHASSIS TEMP,
IZS TEMP
1
,
PMT TEMP,
HVPS VOLTAGE
Diagnostic analog output ID. Enclose
value in double quotes (“) when setting
from the RS-232 interface.
REMOTE_CAL_MODE —
LOW
LOW,
HIGH,
CO2
11
,
O2
10
Range to calibrate during contact-
closure and Hessen calibration.
Enclose value in double quotes (“)
when setting from the RS-232
interface.
PASS_ENABLE —
OFF
OFF,
ON
ON enables passwords; OFF disables
them.
RCELL_CYCLE Seconds
2 0.5–30 Reaction cell temperature control cycle
period.
RCELL_PROP 1/ºC
0.3 (prop.
band = 3.3
ºC)
0–10
Reaction cell temperature PID
proportional coefficient.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...