APPENDIX A - Version Specific Software Documentation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
A-38
05036 Rev D
MODBUS
Register Address
(dec., 0-based)
Description
Units
110
12
SO
2
concentration for range #1, with O
2
correction
PPB
112
12
SO
2
concentration for range #2, with O
2
correction
PPB
200
11
CO
2
concentration
%
202
11
CO
2
concentration during zero/span calibration, just before
computing new slope and offset
%
204
11
CO
2
slope
—
206
11
CO
2
offset
%
MODBUS Floating Point Holding Registers
(32-bit IEEE 754 format; read/write in high-word, low-word order; read/write)
0 Maps
to
SO2_SPAN1 variable; target conc. for range #1
Conc. units
2 Maps
to
SO2_SPAN2 variable; target conc. for range #2
Conc. units
100
10
Maps
to
O2_TARG_SPAN_CONC variable
%
200
11
Maps
to
CO2_TARG_SPAN_CONC variable
%
MODBUS Discrete Input Registers
(single-bit; read-only)
0
PMT detector warning
1
UV detector warning
2
Dark calibration warning
3
Box temperature warning
4
PMT temperature warning
5
Reaction cell temperature warning
6
Sample pressure warning
7 HVPS
warning
8
System reset warning
9
Rear board communication warning
10
Relay board communication warning
11
Front panel communication warning
12
Analog calibration warning
13
Dynamic zero warning
14
Dynamic span warning
15 Invalid
concentration
16
In zero calibration mode
17
In span calibration mode
18
In multi-point calibration mode
19
System is OK (same meaning as SYSTEM_OK I/O signal)
20 Sample
flow
warning
21
1
IZS temperature warning
22
2
In low span calibration mode
23
2
Vacuum pressure warning
24
3
SO
2
concentration alarm limit #1 exceeded
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...