4.8.9. FLOW CALIBRATION
The flow calibration allows the user to adjust the values of the sample flow rates as they
are displayed on the front panel and reported through COM ports to match the actual
flow rate measured at the sample inlet. This does not change the hardware measurement
of the flow sensors, only the software calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the
SAMPLE inlet (refer to Section 11 for more details).
Once the flow meter is attached and is measuring actual gas flow, access the Signal I/O
from the DIAG Menu (refer to Figure 4-17), then press:
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
FLOW CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
ACTUAL FLOW: 607 CC / M
0 6 0 7 ENTR EXIT
E
NTR
accepts the
new value
EXIT
ignores the new
value
Adjust these values until the
displayed flow rate equals the
flow rate being measured by the
independent flow meter.
Figure 4-35:
DIAG
– Flow Calibration
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...