Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
156
5.3.1.2.
CONTROL INPUTS
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes
are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear
panel. These are opto-isolated, digital inputs that are activated when a 5 VDC signal
from the “U” pin is connected to the respective input pin.
Table 5-9: Control Input Pin Assignments
INPUT
STATUS
CONDITION WHEN ENABLED
A
External Zero Cal
Zero calibration mode is activated. The mode field of the display will read
ZERO CAL R.
B
External Span Cal
Span calibration mode is activated. The mode field of the display will read
SPAN CAL R.
C
Unused
D
Unused
E
Unused
F
Unused
Digital Ground
Provided to ground an external device (e.g., recorder).
U
DC Power For Input
Pull Ups
Input for +5 VDC required to activate inputs A - F. This voltage can be taken
from an external source or from the “+” pin.
+
In5v Supply
Internal source of +5V which can be used to activate inputs when connected
to pin U.
There are two methods to activate control inputs. The internal +5V available from the
“+” pin is the most convenient method (Figure 5-24). However, to ensure that these
inputs are truly isolated, a separate, external 5 VDC power supply should be used
(Figure 5-25).
CONTROL IN
A B C D E F U
+
SPAN
ZERO
Figure 5-24:
Control Inputs with Local 5 V Power Supply
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...