
Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
128
5.1.2.1.
CONFIGURING ETHERNET COMMUNICATION MANUALLY (STATIC IP ADDRESS)
1. Connect a cable from the
analyzer’s Ethernet port to a Local Area Network (LAN) or
Internet port.
2.
From the analyzer’s front panel touchscreen, access the Communications Menu as
shown in Figure 5-1.
3. Follow the setup sequence as shown in Figure 5-3, and edit the Instrument and
Gateway IP addresses and Subnet Mask to the desired settings.
4. From the computer, enter the same information through an application such as
HyperTerminal.
Table 5-2 shows the default Ethernet configuration settings.
Table 5-2: LAN/Internet Default Configuration Properties
PROPERTY
DEFAULT STATE
DESCRIPTION
DHCP
ON
This displays whether the DHCP is turned ON or OFF.
Press EDIT and toggle ON for automatic configuration
after first consulting network administrator. (
INSTRUMENT
IP ADDRESS
This string of four packets of 1 to 3 numbers each (e.g.
192.168.76.55.) is the address of the analyzer itself.
GATEWAY IP
ADDRESS
0.0.0.0
Can only be edited when DHCP is set to OFF.
A string of numbers very similar to the Instrument IP
address (e.g. 192.168.76.1.) that is the address of the
computer used by your LAN to access the Internet.
SUBNET MASK
0.0.0.0
Can only be edited when DHCP is set to OFF.
Also a string of four packets of 1 to 3 numbers each (e.g.
255.255.252.0) that identifies the LAN to which the device
is connected.
All addressable devices and computers on a LAN must
have the same subnet mask. Any transmissions sent to
devices with different subnets are assumed to be outside
of the LAN and are routed through the gateway computer
onto the Internet.
TCP PORT
1
3000
This number defines the terminal control port by which the
instrument is addressed by terminal emulation software,
such as Internet or Teledyne ML
’s APICOM.
HOST NAME
T100
The name by which your analyzer will appear when
addressed from other computers on the LAN or via the
Internet. To change, see Section 5.1.2.3.
1
Do not change the setting for this property unless instructed to by Teledyne ML
’s Customer Service
personnel.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...