Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
180
Step Three:
Perform the calibration or calibration check according to the following
flow chart:
Wait until
STABIL
falls below 0.05
ppb. This may
take several
minutes.
ZERO CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the
main
SAMPLE
display
Pressing
ENTR
changes the calibration of the instrument.
If either the
ZERO
or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
Analyzer enters
ZERO
CAL
mode
.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > ZERO CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB SO2=X.XX
X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST
TST >
CAL
CALZ
CALS
SETUP
SAMPLE
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
Scroll the display to the
STABIL
test function. This
function calculates the stability
of the SO
2
measurements.
The value of
STABIL
may jump
significantly. Wait
until it falls below
0.5
ppb. This may take
several minutes.
Pressing
ENTR
changes the calibration of the instrument.
Analyzer enters
SPAN
CAL
Mode.
SPAN CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
SPAN CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
Figure 6-7: Setup for Manual Calibration with Z/S Valve Option Installed (Step 3)
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...