Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
280
06807A(DCNxxxx)
10.6.10. SHUTTER SYSTEM
To check the functionality of the UV light Shutter by manually activating it:
EXIT
4x’s to return
to the
SAMPLE
display
DIAG I / O
33) UVLAMP_SIGNAL= 3.4 MV
PREV NEXT
JUMP
PRNT
EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
PREV NEXT JUMP
ENTR
EXIT
DIAG I / O
0)
EXT_ZERO_CAL=OFF
PREV NEXT
JUMP
PRNT EXIT
Toggle these keys
until
29
is displayed
DIAG I / O
JUMP TO: 01
2 9
ENTR
EXIT
Activate the
Dark Shutter
DIAG I / O
29DARK_SHUTTER=OFF
PREV NEXT
JUMP
OFF
PRNT
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
EXIT
returns
to the main
SAMPLE
display
DIAG I / O
29DARK_SHUTTER=ON
PREV NEXT
JUMP
ON
PRNT
EXIT
DIAG I / O
JUMP TO: 01
33
ENTR
EXIT
Toggle these keys
until
33
is displayed
UV LAMP_SIGNAL
should be
<20 mV
Figure 10-5:
Manual Activation of the UV Light Shutter
10.6.11. PMT SENSOR
The photo multiplier tube detects the light emitted by the UV excited fluorescence of SO
2
.
It has a gain of about 500000 to 1000000. It is not possible to test the detector outside of
the instrument in the field. The best way to determine if the PMT is working properly is
by using the optical test (
OTEST
), which is described in Section 4.8.5. The basic method
to diagnose a PMT fault is to eliminate the other components using
ETEST
,
OTEST
and
specific tests for other sub-assemblies.
10.6.12. PMT PREAMPLIFIER BOARD
To check the correct operation of the preamplifier board, we suggest the technician carry
out the electrical and optical tests described in 4.8.5 and 4.8.6.
If the
ETEST
fails, the preamplifier board may be faulty.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...