
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Troubleshooting & Repair
06807A(DCNxxxx)
293
1. Set the analyzer display to show the signal I/O function,
UVLAMP_SIGNAL
(refer to
Section 10.1.3).
UVLAMP_SIGNAL
is function 33.
2. Slightly loosen the large brass thumbscrew located on the shutter housing (refer to
Figure 10-14) so that the lamp can be moved.
DO NOT
Use Lamp Cap to
adjust Lamp position
UV Lamp
Power Supply
Wires
Thumb screw
UV Lamp
Bracket
Mounting
Screws
Adjust Lamp
Position by
grasping Lamp
body ONLY
Figure 10-14. UV Lamp Adjustment
ATTENTION
COULD DAMAGE INSTRUMENT
DO NOT grasp the UV lamp by its cap when changing its position - always grasp the main
body of the lamp (refer to Figure 10-13) Inattention to this detail could twist and potentially
disconnect the lamp’s power supply wires.
3. While watching the
UVLAMP_SIGNAL
reading, slowly rotate the lamp or move it
back and forth vertically until the
UVLAMP_SIGNAL
reading is at its maximum.
4. Compare the
UVLAMP_SIGNAL
reading to the information in Table 10-10 and follow
the instructions there.
Table 10-10:
Example of HVPS Power Supply Outputs
UVLAMP_SIGNAL
ACTION TO BE TAKEN
3500mV±200mV.
No Action Required
>
4900mV at any time.
Adjust the UV reference detector potentiometer (Figure 10-15) until
UVLAMP_SIGNAL
reads approximately 3600mV before continuing to adjust the
lamp position.
>3700mV or < 3300mV
Adjust the UV reference detector potentiometer (Figure 10-15) until
UVLAMP_SIGNAL
reads as close to 3500mV as possible.
.< 600mV
Replace the lamp (Section 10.7.2.6.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...