Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
292
06807A(DCNxxxx)
output.
Lamp Positioning
– The UV output level of the lamp is not even across the entire length
of the lamp. Some portions of the lamp shine slightly more brightly than others. At the
factory the position of the UV lamp is adjusted to optimize the amount of UV light
shining through the UV filter/lens and into the reaction cell. Changes to the physical
alignment of the lamp can affect the analyzers ability to accurately measure SO
2
.
Figure 10-13: Shutter Assembly
CAUTION
GENERAL SAFETY HAZARD
ALWAYS wear UV-Protective, Safety Glasses when working with the UV Lamp
Assembly.