
Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
260
06807A(DCNxxxx)
A systematic approach to troubleshooting will generally consist of the following five
steps:
1. Note any
WARNING MESSAGES
and take corrective action as necessary.
2. Examine the values of all TEST functions and compare them to factory values. Note
any major deviations from the factory values and take corrective action.
3. Use the internal electronic status LEDs to determine whether the electronic
communication channels are operating properly.
Verify that the DC power supplies are operating properly by checking the voltage test points
on the relay PCA.
Note that the analyzer’s DC power wiring is color-coded and these colors match the color of
the corresponding test points on the relay PCA.
4. Suspect a leak first!
Customer service data indicate that the majority of all problems are eventually traced to leaks
in the internal pneumatics of the analyzer or the diluent gas and source gases delivery
systems.
Check for gas flow problems such as clogged or blocked internal/external gas lines, damaged
seals, punctured gas lines, a damaged / malfunctioning pumps, etc.
5. Follow the procedures defined in Section 10.6
to confirm that the analyzer’s vital
functions are working (power supplies, CPU, relay PCA, touch-screen display, PMT
cooler, etc.).
Refer to Figure 3-5 for the general layout of components and sub-assemblies in the analyzer.
Refer to the wiring interconnect diagram and interconnect list in Appendix D.
10.1.1. FAULT DIAGNOSTICS WITH WARNING MESSAGES
The most common and/or serious instrument failures will result in a warning message
displayed on the front panel. Table
10-1 contains a list of warning messages, along with
their meaning and recommended corrective action.
It should be noted that if more than two or three warning messages occur at the same
time, it is often an indication that some fundamental analyzer sub-system (power supply,
relay board, motherboard) has failed rather than an indication of the specific failures
referenced by the warnings. In this case, a combined-error analysis needs to be
performed.
The analyzer will alert the user that a Warning message is active by flashing the FAULT
LED and displaying the Warning message in the Param field along with the
CLR
button
(press to clear Warning message). The
MSG
button displays if there is more than one
warning in queue or if you are in the TEST menu and have not yet cleared the message.
The following display/touchscreen examples provide an illustration of each:
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...