Theory Of Operation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
240
06807A(DCNxxxx)
The motherboard is directly mounted to the rear panel and collects data, performs signal
conditioning duties and routes incoming and outgoing signals between the CPU and the
analyzer’s other major components.
Concentration data of the T100 are generated by the Photo Multiplier Tube (PMT),
which produces an analog current signal corresponding to the brightness of the
fluorescence reaction in the sample chamber. This current signal is amplified to a DC
voltage signal (front panel test parameter
PMT
) by a PMT preamplifier printed circuit
assembly (located on top of the sensor housing).
PMT
is converted to digital data by a
bi-polar, analog-to-digital converter, located on the motherboard.
In addition to the PMT signal, a variety of sensors report the physical and operational
status of the analyzer’s major components, again through the signal processing
capabilities of the motherboard. These status reports are used as data for the SO
2
concentration calculation (e.g. pressure and temperature reading used by the
temperature/pressure compensation feature) and as trigger events for certain warning
messages and control commands issued by the CPU. They are stored in the CPU’s
memory and, in most cases, can be viewed through the front panel display.
The CPU communicates with the user and the outside world in a variety of ways:
Through the analyzer’s front panel LCD touch-screen interface
RS-232 and RS-485 serial I/O channels
Various analog voltage and current outputs
Several digital I/O channels
Ethernet
Finally, the CPU issues commands (also over the I
2
C bus) to a series of relays and
switches located on a separate printed circuit assembly, the relay board (located in the
rear of the chassis on its own mounting bracket) to control the function of key
electromechanical devices such as valves and heaters.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...