Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
65
3.5.6. PNEUMATIC LAYOUT WITH CO
2
SENSOR OPTION
Figure 3-22 shows the internal, pneumatic connections for the analyzer with the carbon
dioxide (CO
2
) sensor option installed. Pneumatically, the CO
2
sensor is placed in line
with the sample gas line between the particulate filter and the analyzer’s sample
chamber. It does not alter the gas flow rate of the sample through the analyzer.
FLOW PRESSURE
SENSOR PCA
Chassis
with CO
2
Sensor Option
EXHAUST
GAS OUTLET
UV
LAMP
PMT
REACTION
CELL
PUMP
Particulate
Filter
HYDROCARBON
SCRUBBER
(Kicker)
V
A
C
U
U
M
M
A
N
IF
O
L
D
CO
2
Probe
SAMPLE
GAS INLET
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Figure 3-22:
Pneumatic Layout with CO
2
Sensor
3.5.7. CALIBRATION GASES
Zero air and span gas are required for accurate calibration.
3.5.7.1.
ZERO AIR
A gas that is similar in chemical composition to the earth’s atmosphere but without the
gas being measured by the analyzer, in this case SO
2
. If your analyzer is equipped with
an Internal Zero Span (IZS) or an external zero air scrubber option, it is capable of
creating zero air.
For analyzers without an IZS or external zero air scrubber option, a zero air generator
such as the Teledyne Model 701 can be used (Figure 3-16).
3.5.7.2.
SPAN GAS
A gas specifically mixed to match the chemical composition of the type of gas being
measured at near full scale of the desired measurement range. In this case, SO
2
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...