Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Specifications, Approvals & Compliance
31
Table 2-3: CO
2
Sensor Option Specifications
2.2. EPA EQUIVALENCY DESIGNATION
The T100 Analyzer is designated as Reference Method Number EQSA-0495-100 as per
40 CFR Part 53 when operated under the following conditions:
Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
Ambient temperature range of 5
o
C to 40
o
C.
Line voltage range of 100-120 VAC or 220-240 VAC, at 50 or 60 Hz.
Sample filter: Equipped with PTFE filter element in the internal filter assembly.
Sample flow of 650 +/- 65 cm
3
/min.
Vacuum pump (internal) capable of 14"Hg Absolute pressure @ 1 slpm or better.
Software settings:
Dynamic span
OFF
Dynamic zero
OFF
Dilution factor
OFF
AutoCal
ON or OFF
Dual range
ON or OFF
Auto-range
ON or OFF
Temp/Pressure compensation
ON
Under this designation, the analyzer may be operated with or without the following
optional equipment:
Rack mount with chassis slides.
Rack mount without slides, ears only.
Zero/span valve options.
Internal zero/span (IZS) option with either:
SO
2
permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.
SO
2
permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.
Under the designation, the IZS option cannot be used as the source of
calibration.
Span Drift (7 days)
<± 0.1% CO
2
Accuracy
<± (0.02% CO
2
+ 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...