Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
290
06807A(DCNxxxx)
8. Using a lint-free cloth dampened with distilled water, clean the lens, the filter and all
of the housing assembly mechanical parts.
9. Dry everything with a 2nd lint-free cloth.
10. Reassemble the lens/filter housing (refer to Figure 10-11 and Figure 10-10).
IMPORTANT
Use gloves and a clean plastic covered surface during assembly. Cleanliness of the
inside of the light shield, the UV lens filter housing and the PMT lens is especially
important.
Note
Ensure to apply loctite to the four lens holder screws and the two light shield screws.
11. Reattach the lens / filter housing to the sample chamber.
12. Reattach the sample chamber to the PMT housing.
13. Reinstall the sensor module into the T100.
14. Close the instrument.
15. Turn the T100 on and let it warm up for 60 minutes.
16. Perform a leak check (refer to Section 8.3.6).
17. Calibrate the analyzer (refer to Section 6).
10.7.2.4.
REPLACING THE UV FILTER/LENS
IMPORTANT
Be careful not to leave thumbprints on the interior of the sample chamber. The
various oils that make up fingerprints fluoresce brightly under UV light and will
significantly a
ffect the accuracy of the analyzer’s SO
2
measurement).
To Replace the UV Filter Lense:
1.
Turn off the instrument’s power and remove the power cord from the instrument.
2. Unplug J4 connector from the motherboard to allow tool access.
3. Alternatively, remove the sensor module as described in Section 10.7.2.1.
4. Remove 4 screws from the shutter cover (refer to Figure 10-13) and remove the
cover.
5. Remove 4 screws from the UV filter retainer.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...