Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
A-37
APPENDIX A-7: MODBUS Register Map
MODBUS
Register Address
(dec., 0-based)
Description
Units
MODBUS Floating Point Input Registers
(32-bit IEEE 754 format; read in high-word, low-word order; read-only)
0
PMT detector reading
mV
2
UV lamp intensity reading
mV
4
UV lamp ratio of calibrated intensity
%
6
PMT electrical offset
mV
8
UV lamp electrical offset
mV
10 SO
2
slope for range #1
—
12 SO
2
slope for range #2
—
14 SO
2
offset for range #1
mV
16 SO
2
offset for range #2
mV
18 SO
2
concentration for range #1 during zero/span
calibration, just before computing new slope and offset
PPB,
PPM
2
20 SO
2
concentration for range #2 during zero/span
calibration, just before computing new slope and offset
PPB
22 SO
2
concentration for range #1
PPB
24 SO
2
concentration for range #2
PPB
26 Concentration
stability
PPB
28 Stray
light
reading
PPB
30 Reaction
cell
temperature
C
32 PMT
temperature
C
34 Sample
pressure
“Hg
36 Internal
box
temperature
C
38
High voltage power supply output
Volts
40
Diagnostic test input (TEST_INPUT_8)
mV
42
Diagnostic temperature input (TEMP_INPUT_5)
C
44
Diagnostic temperature input (TEMP_INPUT_6)
C
46
Ground reference (REF_GND)
mV
48
4096 mV reference (REF_4096_MV)
mV
50 Sample
flow
cc/m
52
1
IZS
temperature
C
54
2
Vacuum
pressure
“Hg
56
1
Pre-amplified UV lamp intensity reading
mV
100
10
O
2
concentration
%
102
10
O
2
concentration during zero/span calibration, just before
computing new slope and offset
%
104
10
O
2
slope
—
106
10
O
2
offset
%
108
10
O
2
sensor cell temperature
C
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...