APPENDIX A - Version Specific Software Documentation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
A-32
05036 Rev D
Table A-6: DAS Parameters, Software Version G.4
NAME
DESCRIPTION
UNITS
PMTDET
PMT detector reading
mV
PHABS
1
Pre-amplified UV lamp intensity reading
mV
UVDET
UV lamp intensity reading
mV
LAMPR
UV lamp ratio of calibrated intensity
%
DRKPMT PMT
electrical
offset
mV
DARKUV
UV lamp electrical offset
mV
SLOPE1 SO
2
slope for range #1
—
SLOPE2 SO
2
slope for range #2
—
OFSET1 SO
2
offset for range #1
mV
OFSET2 SO
2
offset for range #2
mV
CO2SLP
11
CO
2
slope
—
CO2OFS
11
CO
2
offset
%
O2SLPE
10
O
2
slope
—
O2OFST
10
O
2
offset
%
ZSCNC1
SO
2
concentration for range #1 during zero/span calibration, just before
computing new slope and offset
PPB,
PPM
2
ZSCNC2
SO
2
concentration for range #2 during zero/span calibration, just before
computing new slope and offset
PPB
CO2ZSC
11
CO
2
concentration during zero/span calibration, just before computing new
slope and offset
%
O2ZSCN
10
O
2
concentration during zero/span calibration, just before computing new
slope and offset
%
CONC1 SO
2
concentration for range #1
PPB
CONC2 SO
2
concentration for range #2
PPB
CNCBK1
4
SO
2
concentration plus background concentration for range #1
PPB
CNCBK2
4
SO
2
concentration plus background concentration for range #2
PPB
BKGND1
4
Background
concentration for range #1
PPB
BKGND2
4
Background
concentration for range #2
PPB
SO2CR1
12
SO
2
concentration for range #1, with O
2
correction
PPB
SO2CR2
12
SO
2
concentration for range #2, with O
2
correction
PPB
CO2CNC
11
CO
2
concentration
%
O2CONC
10
O
2
concentration
%
STABIL Concentration
stability
#1
PPB
STABL2
3
Concentration
stability
#2
PPB
STABUV
3
UV lamp stability
mV
STRLGT
Stray light reading
PPB
RCTEMP
Reaction cell temperature
C
O2TEMP
10
O
2
sensor cell temperature
C
IZSTMP
1
IZS
temperature
C
PMTTMP PMT
temperature
C
SMPFLW Sample
flow
cc/m
SMPPRS Sample
pressure
“Hg
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...