Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
83
4.2.2. WARNING MESSAGES
The most common instrument failures will be reported as a warning on the analyzer’s
front panel and through the COMM ports. Section 10.1.1 explains how to use these
messages to troubleshoot problems. Section 10.1.3 shows how to view and clear
warning messages. Table 4-3 lists all warning messages for the current version of
software.
Table 4-3: List of Warning Messages
MESSAGE
MEANING
ANALOG CAL WARNING
The instrument's A/D circuitry or one of its analog outputs is not calibrated.
BOX TEMP WARNING
The temperature inside the T100 chassis is outside the specified limits.
CANNOT DYN SPAN
Remote span calibration failed while the dynamic span feature was set to turned on
CANNOT DYN ZERO
Remote zero calibration failed while the dynamic zero feature was set to turned on
CONFIG INITIALIZED
Configuration was reset to factory defaults or was erased.
DARK CAL WARNING
Dark offset above limit specified indicating that too much stray light is present in the
sample chamber.
DATA INITIALIZED
DAS data storage was erased.
HVPS WARNING
High voltage power supply for the PMT is outside of specified limits.
IZS TEMP WARNING
On units with IZS options installed: The permeation tube temperature is outside of
specified limits.
PMT DET WARNING
PMT detector output outside of operational limits.
PMT TEMP WARNING
PMT temperature is outside of specified limits.
RCELL TEMP WARNING
Sample chamber temperature is outside of specified limits.
REAR BOARD NOT DET
The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN
The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN
The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN
Sample pressure outside of operational parameters.
SYSTEM RESET
The computer was rebooted.
UV LAMP WARNING
The UV lamp intensity measured by the reference detector reading too low or too
high
To view and clear warning messages, press:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
TEST
CAL
MSG
CLR
SETUP
Make sure warning messages are
not due to real problems.
Press
CLR
to clear the current
message.
If more than one warning is active, the
next message will take its place
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
mode.
SAMPLE
RANGE=500.000 PPM
SO2=XXX.X
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
HVPS WARNING
SO2=XXX.X
TEST CAL
MSG
CLR
SETUP
TEST
ignores warning messages
MSG
activates warning
messages.
<TST TST
> buttons replaced with
TEST
button
All Warning messages are hidden,
but
MSG
button appears
NOTE:
If the warning message persists
after several attempts to clear it,
the message may indicate a real
problem and not an artifact of the
warm-up period.
Figure 4-3: Viewing and Clearing T100 WARNING Messages
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...