Operating Instructions
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
94
4.6.4. SINGLE RANGE MODE (SNGL)
The default range mode for the analyzer is single range, in which all analog
concentration outputs are set to the same reporting range. This reporting range can be set
to any value between 0.1 ppb and 20,000 ppb.
While the two outputs always have the same reporting range, the span and scaling of
their electronic signals may also be configured for different differently (e.g.,
A1
= 0-10
V;
A2
= 0-0.1 V).
To select
SNGLE
range mode and to set the upper limit of the range, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
RANGE:
500.0 Conc
0 0
5
0 0 .0
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
Figure 4-11:
SETUP RNGE
– Single Range Mode
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...