Remote Control Commands
R&S
®
SMBV100B
830
User Manual 1178.4460.02 ─ 03
Manual operation:
See
[:SOURce]:INPut:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Example:
INP:TRIG:SLOP NEG
Activates the falling slope of the external trigger signal at the
trigger input.
Manual operation:
See
12.16.8
SOURce:IQ Subsystem
The
SOURce:IQ
subsystem contains the commands for configuring the I/Q modula-
tion.
Required options
See
Chapter 5.1, "Required Options"
...........................................................................................830
.............................................................................................. 831
................................................................................................831
............................................................................................ 831
......................................................................................831
.................................................................................. 831
.......................................................................................... 832
[:SOURce<hw>]:IQ:SOURce
<Source>
Selects the input signal source for the I/Q modulator.
Parameters:
<Source>
BASeband | ANALog
External signals disable the amplitude modulation, an enabled
custom digital modulation, any configured digital standard or an
applied digital baseband input signal.
Example:
SOURce1:IQ:SOURce ANALog
Selects an external analog signal as the input signal. The signal
must be applied at the I and Q inputs.
Manual operation:
See
SOURce Subsystem