Configuring the RF Signal
R&S
®
SMBV100B
304
User Manual 1178.4460.02 ─ 03
2. Select "Reference Output".
In the "Reference Output" tab, you can set the reference frequency value at the
output connectors.
As a result of parameter dependencies, "Preset This Parameter" sometimes does
not affect output dialogs.
The remote commands required to define these settings are described in
ter 12.16.16, "SOURce:ROSCillator Subsystem"
Settings:
Reference Output/1GHz Reference Output
................................................................ 304
Reference Output/1GHz Reference Output
Selects the signal that is to be output as frequency reference for downstream instru-
ments.
"Off"
Disables the reference signal output.
"10 MHz"
Derives a signal with 10 MHz frequency from the internal reference
oscillator and forwards this signal to the output.
"1 GHz"
Option: R&S
SMBVB-B3
Outputs a 1 GHz signal.
"Input Signal (loop through)"
Forwards the external reference frequency to the output directly.
Remote command:
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
6.8.5
Adjustment Settings
Access:
1. Select "RF" > "Reference Frequency"
2. Select "Adjustment".
Reference Oscillator